Sequence:end, Sequence:point:resistance – GW Instek PEL-2000 Series Programming User Manual User Manual
Page 150

PEL-2000 Programming Manual
150
Query Example
:SEQ:EDIT:POIN?
3
The current point is 3.
:SEQuence:END
Channel Specific
Command
Description
Sets the number of points in the sequence.
Syntax
:SEQuence:END <NR1>
Parameter
<NR1>
Points
1~120
1~120.
Example
:SEQ:END 5
Sets the max number of
points to 5.
Query Syntax
:SEQuence:END?
Return Parameter <NR1>
Points
1~120
1~120
Query Example
:SEQ:END?
5
There are 5 points in the
sequence.
:SEQuence:POINt:RESistance
Channel Specific
Command
Description
Sets the resistance value for the current point. CR
mode only.
Syntax
:SEQuence:POINt:RESistance
<NRf>[OHM]|MIN|MAX
Parameter
<NRf>[OHM], MIN, MAX Resistance value
100
100Ω
100 OHM
100Ω
MAX/MIN
Maximum or minimum
value.
Example
:SEQ:POIN:RES 100
Sets the resistance to 100.
Query Syntax
:SEQuence:POINt::RESistance? [MAX|MIN]
Return Parameter <NR1>
Resistance Value