Power:recall, Power:low:avalue/bvalue – GW Instek PEL-2000 Series Programming User Manual User Manual
Page 137

COMMAND DETAILS
137
Query Example
:POW:CURR?
7.0
Returns the current limit
for the specific channel.
:POWer:RECall
Channel Specific
Command
Description
Sets or queries whether A Value or B Value is the
currently active value in CP mode.
Syntax
:POWer:RECall A/0, B/1
Parameter
A/0, B/1
Value
A/0
A
B/1
B
Example
:POW:REC 1
Makes B Value the active
value.
Query Syntax
:POWer:RECall?
Return Parameter <NR1>
Value
0
A
1
B
Query Example
:POW:REC?
0
A Value is active.
:POWer:LOW:AVALue/BVALue
Channel Specific
Command
Description
Sets the low range A/B Value for constant power
mode.
Syntax
:POWer:LOW:AVALue/BVALue <NRf+>[W]
Parameter
NRf+[W]
AVALue 1
Sets A Value to 1 watt.
BVALue 1W
Sets B Value to 1 watt.
AVALue MIN
Sets A Value to the
minimum level for the
specific channel.