Sequence:repeat, Sequence:end:load – GW Instek PEL-2000 Series Programming User Manual User Manual
Page 153

COMMAND DETAILS
153
Return Parameter <NR2>
Point
0.0001~60000
Returns the point duration
time.
MAX/MIN
Returns the allowable
maximum and minimum.
Query Example
:SEQ:POIN:TIME?
0.00100
The point duration time
is 0.001 seconds.
:SEQuence:REPeat
Channel Specific
Command
Description
Sets the number of times the sequence can be
repeated (looped).
Syntax
:SEQuence:REPeat <NR1>
Parameter
<NR1>
Repeat setting
1~9999
1~9999
0
Infinite repeats
Example
:SEQ:REP 10
Repeat 10 times
:SEQ:REP 0
Repeat infinitely
Query Syntax
:SEQuence:REPeat?
Return Parameter <NR1>
Repeat setting
1~9999
1~9999
0
Infinite
Query Example
:SEQ:REP?
10
Repeats the sequence 10
times
:SEQuence:END:LOAD
Channel Specific
Command
Description
Sets On End Load to On or Off. On End Load
determines if the channel will set the load on or off
at the end of its sequence until the end of the last
sequence (of ALL other channels).
Syntax
:SEQuence:END:LOAD {OFF|0|ON|1}