Simple Repair Procedures

The procedures described in this section can be done with the quaternary
pump in place in the system stack.

Table 7

Simple Repair Procedures

Procedure

Typical Frequency

Notes

“Exchanging the Active Inlet Valve
Cartridge or the Active Inlet Valve” on
page 103

If internally leaking

Pressure ripple unstable, run leak test for
verification

“Exchanging the Outlet Ball Valve” on
page 106

If internally leaking

Pressure ripple unstable, run leak test for
verification

“Exchanging the Purge Valve Frit or the
Purge Valve” on page 108

If internally leaking

Solvent dripping out of waste outlet when
valve closed

“Exchanging the Purge Valve Frit or the
Purge Valve” on page 108

If the frit shows indication
of contamination or
blockage

A pressure drop of > 10 bar across the frit
(5 ml/min H

2

O with purge open) indicates

blockage

“Exchanging the Pump Seals and Seal
Wear-in Procedure” on page 112

If pump performance
indicates seal wear

Leaks at lower pump head side, unstable
retention times, pressure ripple unstable —

run leak test for verification

“Exchanging the Plungers” on page 115

If scratched

Seal life time shorter than normally
expected — check plungers while changing
the seals

“Installing the Continuous Seal Wash
Option” on page 116

If seals show indication of
leaks

Leaks at lower pump head side, loss of wash
solvent