185

Parts and Materials

Remote Cables

Remote Cables

One end of these cables provides a Agilent Technologies APG (Analytical
Products Group) remote connector to be connected to Agilent 1100 Series
modules. The other end depends on the instrument to be connected to.

Agilent 1100 to 3390 Integrators

Connector
01046-60203

Pin
3390

Pin
Agilent 1100

Signal Name

Active
(TTL)

2

1 - White

Digital ground

NC

2 - Brown

Prepare run

Low

7

3 - Gray

Start

Low

NC

4 - Blue

Shut down

Low

NC

5 - Pink

Not connected

NC

6 - Yellow

Power on

High

NC

7 - Red

Ready

High

NC

8 - Green

Stop

Low

NC

9 - Black

Start request

Low