Encoder missing – Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 70

Advertising
background image

70

Troubleshooting and Test Functions

Encoder Missing

Encoder Missing

The optical encoder on the pump motor in the quaternary pump is missing or
defective.

The processor checks the presence of the pump encoder connector every
2 seconds. If the connector is not detected by the processor, the error
message is generated.

Probable Causes

Defective or disconnected pump encoder connector.

Defective pump drive assembly.

Suggested Actions

❏ Ensure the connector is clean, and seated correctly.

❏ Exchange the pump drive assembly.

Advertising