Temperature limit exceeded – Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 73

Advertising
background image

73

Troubleshooting and Test Functions

Temperature Limit Exceeded

Temperature Limit Exceeded

The temperature of one of the motor-drive circuits is too high.

The processor continually monitors the temperature of the drive circuits on
the LPM board. If excessive current is being drawn for long periods, the
temperature of the circuits increases. If the temperature exceeds the upper
limit of 95 ºC, the error message is generated.

Probable Causes

High friction (partial mechanical blockage) in the pump drive assembly.

Partial blockage of the flowpath in front of the damper.

Defective pump drive assembly.

Defective LPM board.

Suggested Actions

❏ Ensure the capillaries and frits between the pump head and damper inlet

are free from blockage.

❏ Ensure the outlet valve is not blocked.

❏ Remove the pump head assembly. Ensure there is no mechanical blockage

of the pump head assembly or pump drive assembly.

❏ Exchange the pump drive assembly.

❏ Exchange the LPM board.

Advertising