Introduction to the Quaternary Pump

Overview

Figure 42

Principle of the Quaternary Pump

When turned on, the quaternary pump runs through an initialization
procedure to determine the upper dead center of the first plunger. The first
plunger moves slowly upwards into the mechanical stop of chamber and
from there it moves back a predetermined path length. The controller stores
this plunger position in memory. After this initialization the quaternary pump
starts operation with the set parameters. The active inlet valve is opened and
the down-moving plunger draws solvent into the first chamber. At the same
time the second plunger is moving upwards delivering into the system. After
a controller-defined stroke length (depending on the flow rate) the drive
motor is stopped and the active inlet valve is closed. The motor direction is
reversed and moves the first plunger up until it reaches the stored upper limit
and at the same time moving the second plunger downwards. Then the
sequence starts again moving the plungers up and down between the two
limits. During the up movement of the first plunger the solvent in the

From solvent bottle

Inlet
valve

Outlet
valve

Damper

To mixing chamber

Seal

Chamber 1

Ball screw drive

Motor with encoder

Gear

To waste

Purge valve

Plunger 2

Chamber 2

Plunger 1