Repairing the Pump

Exchanging the Low Pressure Pump Main Board (LPM Board)

print sendmodule$ (lpmp, "ser?") The reply line will give the serial number of
the quaternary pump.

Replacing the Quaternary Pump’s Firmware

The installation of new firmware is required

if a new version solves problems of the currently installed version.

if the version of firmware on the new main board (LPM) after an exchange
of the board is older than the one previously installed.

To upgrade the quaternary pump’s firmware follow the procedures and
instructions given on the internet @:

http://www.agilent.com/chem

to download and install always the newest available version of firmware on
your system or call your local service provider for assistance.