Exchanging the plungers – Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 115

Advertising
background image

115

Repairing the Pump

Exchanging the Plungers

Exchanging the Plungers

When required:

❏When scratched

Tools required:

❏4-mm hexagonal key

Parts:

❏Plunger 5063-6586

1

Disassemble the pump head assembly (see
“Removing and Disassembling the Pump
Head Assembly” on page 110
).

2

Check the plunger surface and remove any
deposits or layers. Cleaning can be done with
alcohol or tooth paste. Replace plunger if
scratched.

3

Reassemble the pump head assembly (see
“Reassembling the Pump Head Assembly” on
page 121
)

Plunger surface