Repairing the Pump

Exchanging the Purge Valve Frit or the Purge Valve

6

Place a new frit into the purge valve with the orientation of the frit as shown
above.

7

Reinstall the cap and the gold seal.

N O T E

Before reinstallation always check the gold seal. A deformed seal should be
exchanged.

8

Insert the purge valve into the pump head and locate the pump outlet capillary
and the waste tube as shown in Figure 18.

9

Tighten the purge valve and reconnect outlet capillary and waste tubing.

Figure 18

Exchanging the Purge Valve