Removing the top cover and foam – Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 128

Advertising
background image

128

Repairing the Pump

Removing the Top Cover and Foam

Removing the Top Cover and Foam

Tools required

Screwdriver Pozidriv #1

Preparations for this
procedure

Switch off quaternary pump at the main power switch.
Disconnect the solvent inlet and outlet tubes from the pump.
Beware of leaking solvents due to hydrostatic flow.
Remove the solvent cabinet from the quaternary pump.
Remove leak funnel with the waste tube from the top cover of the quaternary pump.

1

Remove the front cover by pressing the both
clip fasteners on both sides of the cover.

2

Unplug the power cord and move the lever
towards the power socket.

(1)

Clip fastener (1)

(2)

Lever