52

Troubleshooting and Test Functions

Leak

Leak

A leak was detected in the quaternary pump.

The signals from the two temperature sensors (leak sensor and
board-mounted temperature-compensation sensor) are used by the leak
algorithm to determine whether a leak is present. When a leak occurs, the
leak sensor is cooled by the solvent. This changes the resistance of the leak
sensor which is sensed by the leak-sensor circuit on the LPM board

Probable Causes

Loose fittings.

Broken capillary.

Loose or leaking purge valve, active inlet valve, or outlet ball valve.

Defective pump seals.

Suggested Actions

❏ Ensure all fittings are tight.

❏ Exchange defective capillaries.

❏ Ensure pump components are seated correctly. If there are still signs of a

leak, exchange the appropriate seal (purge valve, active inlet valve, outlet
ball valve).

❏ Exchange the pump seals.