Repairing the Pump

Reassembling the Pump Head Assembly

Reassembling the Pump Head Assembly

Tools required:

❏4-mm hexagonal key

❏Pump head grease (79846-65501)

1

Place the support rings on the plunger
housing (plungers not installed) and snap the
pump head and plunger housing together.

2

Insert the plungers and carefully press them
into the seal.

3

Tighten the lock screw.

Pump head

Support ring

Plunger housing

Plunger

Lock screw