Control Module Screens for the Quaternary Pump

Screens available from the Records screen

Use the Esc key to receive Views on the F5 key. Choose System from the
pull-down menu. Use the F3 key (Records) to select the Generic module. In
this screen the resident firmware revision is shown.

Use the F5 key (FW Update) to enter the Update section. Select the a file
from the PCMCIA card (1311nnnn.DLB) and press execute. When the update
has finished, press F7 key (Transfer) to return the module into the normal
mode (LED on module should stay yellow).

If you have not saved your methods, please do it before continuing.
Otherwise they will be overwritten during the update process.