126

Repairing the Pump

Exchanging the optional Interface Board

Exchanging the optional Interface Board

C A U T I O N

The interface board is sensitive to electrostatic discharge. Always use the ESD
kit when handling electronic boards.

1

Switch off the quaternary pump at the main power switch, unplug the pump
from line power.

2

Disconnect cables from the interfaceboard connectors.

3

Loosen the screws. Slide out the interface board from the quaternary pump.

4

Install the new interface board. Secure screws.

5

Reconnect the cables to the board connector.

6

Reconnect the pump to line power.

Figure 19

Exchanging the Interface Board

When required

Board defective

Parts required

BCD (Interface) board (G1351-68701), see “Optional Interface Boards” on page 216

BCD (Interface) board