Introduction to the Quaternary Pump

The Low-Pressure Pump Main Board (LPM)

Fan Drive

The fan speed is controlled by the main processor depending on the internal
heat distribution in the quaternary pump. The fan provides a PWM signal
which is proportional to its speed. This fan status signal is used for
diagnostics.

Electronic Fuses

The valve circuits are electronically fused on the board. Any error on the
board or shortages of the valves will activate the electronic fuses that will
switch off the supply voltage. This prevents the damage to components.

Onboard Battery

An onboard lithium battery buffers the electronic memories when the
quaternary pump is turned off.

For safety information on Lithium batteries see “Lithium Batteries
Information” on page 263.

Interfaces

The quaternary pump provides the following interfaces:

two CAN connectors as interface to other Agilent 1100 Series modules

one GPIB connector as interface to the ChemStation

one RS-232C for as interface to a computer

one REMOTE connector as interface to other Agilent products

one Analog Output for pressure signal output

one optional interface board

Block Diagram Signal Explanations

ϑ

control signal
data/ address

diagnose signals (digital)
diagnose signals (analog)

temperature of hybrid