221

Introduction to the Quaternary Pump

Agilent 1100 Series Interfaces

Table 33

APG Remote Signal Distribution

Pin

Signal

Description

1

DGND

Digital ground

2

PREPARE

(L) Request to prepare for analysis (for example, detector lamp
on). Receiver is any module performing preanalysis activities.

3

START

(L) Request to start run / timetable. Receiver is any module
performing run-time controlled activities.

4

SHUT DOWN

(L) System has serious problem (for example, leak: stops pump).
Receiver is any module that needs to be shutdown to avoid a
safety risk.

5

Not used

6

POWER ON

(H) All modules connected to system are switched on. Receiver is
any module relying on operation of others.

7

READY

(H) System is ready for next analysis. Receiver is any sequence
controller.

8

STOP

(L) Request to reach system ready state as soon as possible (for
example, stop run, abort or finish and stop injection). Receiver is
any module performing run-time controlled activities.

9

START REQUEST (L) Request to start injection cycle. Receiver is the autosampler.