Pressure above upper limit – Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 61

Advertising
background image

61

Troubleshooting and Test Functions

Pressure Above Upper Limit

Pressure Above Upper Limit

The system pressure has exceeded the upper pressure limit.

Probable Causes

Upper pressure limit set too low.

Blockage in the flowpath (after the damper).

Defective damper.

Defective LPM board.

Suggested Actions

❏ Ensure the upper pressure limit is set to a value suitable for the analysis.

❏ Check for blockage in the flowpath. The following components are

particularly subject to blockage:

purge-valve frit.

needle (autosampler),

seat capillary (autosampler),

sample loop (autosampler), and

column frits.

capillaries with low internal diameters (e.g. 0.12mm id).

❏ Exchange the damper.

❏ Exchange the LPM board.