Repairing the Pump

Exchanging the Active Inlet Valve Cartridge or the Active Inlet Valve

Exchanging the Active Inlet Valve
Cartridge or the Active Inlet Valve

Removing the Active Inlet Valve

1

Unplug the active inlet valve cable from the connector.

2

Disconnect the solvent inlet tube at the inlet valve.

3

Using a 14 mm wrench loosen the active inlet valve and remove the valve from
pump head.

Exchanging the Valve Cartridge

1

Using a pair of tweezers remove the valve cartridge from the actuator
assembly.

2

Before inserting the new valve cartridge clean the area in the actuator
assembly. Fill a syringe with alcohol and flush the cartridge area thouroughly.

3

Insert a new valve cartridge into the actuator assembly. Make sure the valve
cartridge is fully inserted into the actuator assembly.

When required

If internally leaking (backflow)

Tools required

Wrench 14 mm
Pair of Tweezers

Material

Active inlet valve G1312-60010 (complete assembly)
Valve cartridge 5062-8562