Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 144

Advertising
background image

144

Repairing the Pump

Exchanging the Pump Drive

2

Remove the tubings from the gradient
valve (MCGV) and remove the valve
cover (see “Exchanging the Multi-Channel
Gradient Valve (MCGV)” on page 123)
.

3

Disconnect all capillaries and tubes from the
pump head and disconnect the active inlet
valve connector.

4

Loosen the screws of the Z-panel and take it
out of the instrument.

.

5

Using a 4-mm hexagonal key stepwise
loosen and remove the two pump head
screws and remove the pump head from the
pump drive.

Cover

Screws

Z-panel

Pump head screws