Pump head missing – Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 75

Advertising
background image

75

Troubleshooting and Test Functions

Pump Head Missing

Pump Head Missing

The pump-head end stop in the quaternary pump was not found.

When the quaternary pump restarts, the metering drive moves forward to the
mechanical end stop. Normally, the end stop is reached within 20 seconds,
indicated by an increase in motor current. If the end point is not found within
20 seconds, the error message is generated.

Probable Causes

Pump head not installed correctly (screws not secured, or pump head not
seated correctly).

Broken plunger.

Suggested Actions

❏ Install the pump head correctly. Ensure nothing (e.g. capillary) is trapped

between the pump head and body.

❏ Exchange the plunger.

Advertising