Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 225

Advertising
background image

225

Introduction to the Quaternary Pump

Setting the 8-bit Configuration Switch

Switches 1 in down and 2 in up position define that the RS-232C parameters
will be changed. Once the change has been completed, the instrument must
be powered up again in order to store the values in the non-volatile memory.

Use the following tables for selecting the setting which you want to use for
RS-232C communication. The number 0 means that the switch is down and 1
means that the switch is up.

One start bit and one stop bit are always used (not selectable).

Per default, the module will turn into 19200 baud, 8 data bit with no parity.

Table 37

Communication Settings for RS-232C Communication

Mode Select 1

2

3

4

5

6

7

8

RS-232

0

1

Baudrate

Data Bits Parity

Table 38

Baudrate Settings

Switches

Baud Rate

Switches

Baud Rate

3

4

5

3

4

5

0

0

0

9600 (default)

1

0

0

9600

0

0

1

1200

1

0

1

14400

0

1

0

2400

1

1

0

19200

0

1

1

4800

1

1

1

38400

Table 39

Data Bit Settings

Switch 6

Data Word Size

0

7 Bit Communication

1

8 Bit Communication