50

Troubleshooting and Test Functions

Remote Timeout

Remote Timeout

A not-ready condition is still present on the remote input .

When an analysis is started, the system expects all not-ready conditions (e.g.
a not-ready condition during detector balance) to switch to run conditions
within one minute of starting the analysis. If a not-ready condition is still
present on the remote line after one minute the error message is generated.

Probable Causes

Not-ready condition in one of the instruments connected to the remote
line.

Defective remote cable.

Defective components in the instrument showing the not-ready condition.

Suggested Actions

❏ Ensure the instrument showing the not-ready condition is installed

correctly, and is set up correctly for analysis.

❏ Exchange the remote cable.

❏ Check the instrument for defects (refer to the instrument’s reference

documentation).