Troubleshooting and Test Functions

Inlet-Valve Fuse

Inlet-Valve Fuse

The active-inlet valve in the quaternary pump has drawn excessive current
causing the inlet-valve electronic fuse to open.

Probable Causes

Defective active inlet valve.

Defective connection cable (front panel to LPM board).

Defective LPM board.

Suggested Actions

❏ Restart the quaternary pump. If the error message appears again, exchange

the active inlet valve.

❏ Exchange the connection cable.

❏ Exchange the LPM board.