Fan failed – Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 57

Advertising
background image

57

Troubleshooting and Test Functions

Fan Failed

Fan Failed

The cooling fan in the quaternary pump has failed.

The hall sensor on the fan shaft is used by the LPM board to monitor the fan
speed. If the fan speed falls below 2 revolutions/second for longer than
5 seconds, the error message is generated.

Probable Causes

Fan cable disconnected.

Defective fan.

Defective LPM board.

Improperly positioned cables or wires obstructing fan blades.

Suggested Actions

❏ Ensure the fan is connected correctly.

❏ Exchange fan.

❏ Exchange the LPM board.

❏ Ensure the fan is not mechanically blocked.