Repairing the Pump

Exchanging the Damper

3

Disconnect the two damper capillaries.

4

Loosen the screws of the Z-panel and take it
out of the instrument.

5

Disconnect the damper connector at the main
board (J24). Lift the damper out of its
position.

6

Place the new damper into the recess and
connect to the main board (J24).

Damper
capillaries

Screws

Z-panel

Damper

Damper
Connector

Damper
Connector

Damper