Repairing the Pump

Exchanging the Purge Valve Frit or the Purge Valve

Exchanging the Purge Valve Frit or the
Purge Valve

1

Using a 1/4 inch wrench disconnect the pump outlet capillary at the purge
valve.

2

Disconnect the waste tube. Beware of leaking solvents due to hydrostatic
pressure.

3

Using the 14 mm wrench unscrew the purge valve and remove it.

4

Remove the plastic cap with the gold seal from the purge valve.

5

Using a pair of tweezers or a toothpick remove the frit

Figure 17

Purge Valve Parts.

When required

Frit – when plunger seals are exchanged or when contaminated or blocked (pressure drop of
> 10 bar across the frit at a flow rate of 5 ml/min of H

2

0 with purge valve opened)

Purge valve – if internally leaking

Tools required

Wrench 1/4 inch
Wrench 14 mm
Pair of tweezers or toothpick

Parts required

PFTE frit (pack of 5) 01018-22707
Purge valve G1311-60009

Valve body

Gold seal

PTFE frit

Plastic cap