Open cover – Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 58

Advertising
background image

58

Troubleshooting and Test Functions

Open Cover

Open Cover

The top foam has been removed.

The sensor on the LPM board detects when the top foam is in place. If the
foam is removed, the fan is switched off, and the error message is generated.

Probable Causes

The top foam was removed during operation.

Foam not activating the sensor.

Sensor defective.

Rear of the module is exposed to strong direct sunlight.

Suggested Actions

❏ Replace the top foam.

❏ Exchange the LPM board.

❏ Ensure that the rear of module is not directly exposed to strong sunlight.