Exchanging the outlet ball valve – Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 106

Advertising
background image

106

Repairing the Pump

Exchanging the Outlet Ball Valve

Exchanging the Outlet Ball Valve

N O T E

Before exchanging the outlet ball valve you can try to clean it in a sonic bath.
Place the valve in upright position (onto the plastic cap) in a small beaker with
alcohol. Place in a sonic bath for 5 – 10 minutes.

1

Using a 1/4 inch wrench disconnect the valve capillary from the outlet ball
valve.

2

Using the 14 mm wrench loosen the valve and remove it from the pump body.

3

Check that the new valve is assembled correctly and that the gold seal is
present (If the gold seal is deformed, it should be replaced).

Figure 15

Outlet Ball Valve Parts

4

Reinstall the outlet ball valve and tighten the valve.

5

Reconnect the valve capillary.

When required

If internally leaking

Tools required

Wrench 1/4 inch
Wrench 14 mm

Parts required

Outlet ball valve G1311-60012

Valve body

Gold seal

Plastic cap