Shut-down – Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 49

Advertising
background image

49

Troubleshooting and Test Functions

Shut-Down

Shut-Down

An external instrument has generated a shut-down signal on the remote line.

The quaternary pump continually monitors the remote input connectors for
status signals. A LOW signal input on pin 4 of the remote connector generates
the error message.

Probable Causes

Leak detected in another module with a CAN connection to the system .

Leak detected in an external instrument with a remote connection to the
system.

Shut-down in an external instrument with a remote connection to the
system.

The degasser failed to generate sufficient vacuum for solvent degassing.

Suggested Actions

❏ Fix the leak in the external instrument before restarting the quaternary

pump.

❏ Check external instruments for a shut-down condition.

❏ Check the vacuum degasser for an error condition. Refer to the Reference

Manual

for the Agilent 1100 Series vacuum degasser.