Parts and Materials

Remote Cables

Agilent 1100 to HP 1090 LC, HP 1040 DAD or Signal Distribution
Module

Agilent 1100 to General Purpose

Connector
01046-60202

Pin
HP 1090

Pin
Agilent 1100 Signal Name

Active
(TTL)

1

1 - White

Digital ground

NC

2 - Brown

Prepare run

Low

4

3 - Gray

Start

Low

7

4 - Blue

Shut down

Low

8

5 - Pink

Not connected

NC

6 - Yellow

Power on

High

3

7 - Red

Ready

High

6

8 - Green

Stop

Low

NC

9 - Black

Start request

Low

5 - Key

Connector
01046-60201

Pin
Universal

Pin
Agilent 1100

Signal Name

Active
(TTL)

1 - White

Digital ground

2 - Brown

Prepare run

Low

3 - Gray

Start

Low

4 - Blue

Shut down

Low

5 - Pink

Not connected

6 - Yellow

Power on

High

7 - Red

Ready

High

8 - Green

Stop

Low

9 - Black

Start request

Low