Repairing the Pump

Exchanging the Low Pressure Pump Main Board (LPM Board)

Table 8

Identifying Connectors on the LPM Board

Figure 21

Removing the Screws at the Back Plane.

4

Remove the connector screws from the GPIB, Remote and the analog
pressure output connector.

5

Remove the board. Place the board on the ESD kit.

Connector

Description

Connector Description

J1

RS232

J17

Encoder

J2

Remote

J19

AIV

J3

Analog Pressure signal

J23

Leak sensor

J4

GPIB

J24

Damper

J5/6

CAN connector

J25

Fan

J9

Power supply

J26

Interface board

J16

Motor J30

MCGV

Screws and Nut

Analog output

Remote

GPIB

Board recess