Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 117

Advertising
background image

117

Repairing the Pump

Installing the Continuous Seal Wash Option

4

Remove the support rings from the plunger
housing and lift the housing away from the
plungers.

5

Check the plunger surface and remove any
deposits or layers. Cleaning can be done with
alcohol or tooth paste. Replace plunger if
scratched.

6

Install the support ring assembly from the
seal wash option kit into the plunger housing.

7

If necessary replace the pump seals (see
page 112 ). Reassemble the pump head
(see page 121 ).

Support ring

Plunger housing

Plunger

Plunger surface

Seal keeper

Seal wash support ring

Plunger housing

Plunger

Seal