Replacing the active inlet valve – Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 104

Advertising
background image

104

Repairing the Pump

Exchanging the Active Inlet Valve Cartridge or the Active Inlet Valve

Figure 13

Active Inlet Valve Assembly

Replacing the Active Inlet Valve

1

Insert the new, or rebuilt valve into the pump head. Using the 14 mm wrench
turn the nut until hand tight.

2

Position the valve that the solvent inlet tube connection points towards the
front.

3

Using the 14 mm wrench tighten the nut by turning the valve in its final
position (not more than a quarter turn). Do not overtighten the valve. The
solvent inlet tube connection should point to the right corner of the pump
head.

4

Reconnect the solvent inlet tube to active inlet valve and the active inlet valve
cable to the connector in the Z-panel.

5

After an exchange of the valve cartridge it may take several ml of pumping
with the solvent used in the current application, before the flow stabilizes at a
%-ripple as low as it used to be when the system was still working properly.

Valve cartridge

Valve body