Repairing the Pump

Exchanging the Low Pressure Pump Main Board (LPM Board)

Exchanging the Low Pressure Pump Main
Board (LPM Board)

1

Turn off the pump, disconnect all cables and remove the pump from the stack.

2

Remove the top cover and foam (see “Removing the Top Cover and Foam” on
page 128)
.

3

Disconnect all connectors from the main board.

N O T E

When removing connectors, counterhold on connector J3 with one hand.

Figure 20

LPM Board Connectors

When required

Board defective

Tools required

Wrench 14 mm
Wrench 7 mm
Wrench 5 mm

Parts required

LPM Board, G1311-66520, exchange part number G1311-69520

S1

J4

J5

J6

J1

J2

J3

J9

J17

J16

J11

J19

J24 J25

J23

J26

J30