Removing and disassembling the pump head assembly – Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 110

Advertising
background image

110

Repairing the Pump

Removing and Disassembling the Pump Head Assembly

Removing and Disassembling the Pump
Head Assembly

WA R NI N G

Never start the pump when the pump head is removed. This may
damage the pump drive.

When required

Exchanging the seals
Exchanging the plungers
Exchanging seals of the seal wash option

Tools required

Wrench 1/4 inch
4-mm hexagonal key

Preparations for this
procedure

Switch off quaternary pump at the main power switch

1

Disconnect all capillaries and tubes from the
pump head and disconnect the active inlet
valve cable.

2

Using a 4-mm hexagonal key stepwise
loosen and remove the two pump head
screws and remove the pump head from the
pump drive.

Pump head screws