Repairing the Pump

Exchanging the Pump Seals and Seal Wear-in Procedure

Exchanging the Pump Seals and Seal
Wear-in Procedure

When required:

❏Seal leaking, if indicated by the results of the

leak test

Tools required:

❏ 4-mm hexagonal key, Wrench 1/4 inch

Parts required:

❏ Seals (pack of 2) 5063-6589 (standard) or

0905-1420 (for normal phase applications)

For the seal wear-in procedure:

❏ Adapter AIV to inlet tube (0100-1847)

❏ Restriction capillary (5022-2159)

1

Disassemble the pump head assembly (see
“Removing and Disassembling the Pump
Head Assembly” on page 110
).

2

Using one of the plungers carefully remove
the seal from the pump head (be careful not
to break the plunger). Remove wear
retainers, if still present.

3

Clean the pump chambers, ensure all particu-
late matter is removed. Best cleaning results
will be achieved by removing all three valves
(see pages 103, 106 and 108) and injecting
solvent into each chamber.