186

Parts and Materials

Remote Cables

Agilent 1100 to 3392/3 Integrators

Agilent 1100 to 3394 Integrators

N O T E

START and STOP are connected via diodes to pin 3 of the 3394 connector.

Connector
01046-60206

Pin
3392/3

Pin
Agilent 1100

Signal Name

Active
(TTL)

3

1 - White

Digital ground

NC

2 - Brown

Prepare run

Low

11

3 - Gray

Start

Low

NC

4 - Blue

Shut down

Low

NC

5 - Pink

Not connected

NC

6 - Yellow

Power on

High

9

7 - Red

Ready

High

1

8 - Green

Stop

Low

NC

9 - Black

Start request

Low

4 - Key

Connector
01046-60210

Pin
3394

Pin
Agilent 1100

Signal Name

Active
(TTL)

9

1 - White

Digital ground

NC

2 - Brown

Prepare run

Low

3

3 - Gray

Start

Low

NC

4 - Blue

Shut down

Low

NC

5 - Pink

Not connected

NC

6 - Yellow

Power on

High

5,14

7 - Red

Ready

High

6

8 - Green

Stop

Low

1

9 - Black

Start request

Low

13, 15

Not connected

Advertising