Repairing the Pump

Exchanging the Low Pressure Pump Main Board (LPM Board)

8

Press the Execute (F8) key. Below the box, a reply line should then say:
Reply RA 0000 TYPE "G1311A" .

9

Turn off the module, then turn it on again. Turn on should be normal. In the
Records

screen, the product# column should indicate the quaternary pump. If

a chemstation is also connected, re-boot it now.

Entering the Type Command using the ChemStation

The TYPE is entered by typing a specific command into the command line at
the bottom of the main user interface screen.

1

To enter the TYPE for a specific module, type the following command into the
command line:

print sendmodule$(lpmp, "TYPE G1311A")

N O T E

It is important to enter the TYPE command correctly. An incorrect type
command may cause the module to turn on in it’s resident mode. In such a
case, re-enter the TYPE command correctly.

2

The reply line will respond with RA 0000 Type "G1311A".

3

Turn the quaternary pump off, then on again. Then, re-boot the ChemStation.
Boot up and subsequent control of the system should be normal.

4

The TYPE of a module can also be identified by typing the following command
into the command line:

print sendmodule$ (lpmp, "TYPE?")
The reply line will give the module TYPE.

Entering the Serial Number

Entering the Serial Number using the Control Module

1

Connect the control module to the quaternary pump. Turn on the pump.

2

In the control module, press Views (F5) and select the system screen, then
press Records (F4). Using the up/down arrows, make sure that the quaternary
pump is highlighted.

3

Press FW Update (F5). Now, press the m key. This will display a box which
says Update Enter Serial#.