Installing the Pump

Priming and Purging the System

Table 5

Choice of Priming Solvents for Different Purposes

Activity

Solvent

Comments

After an installation

When switching between
reverse phase and normal
phase (both times)

Isopropanol

Isopropanol

Best solvent to flush air out of
the system

Best solvent to flush air out of
the system

After an installation

Ethanol or Methanol

Alternative to Isopropanol
(second choice) if no
Isopropanol is available

To clean the system when
using buffers

After a solvent change

Bidistilled water

Bidistilled water

Best solvent to re-dissolve
buffer cristals

Best solvent to re-dissolve
buffer cristals

After the installation of
normal phase seals (P/N
0905-1420)

Hexane + 5% Isopropanol

Good wetting properties