Potential causes of pressure test failure – Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 87

Advertising
background image

87

Troubleshooting and Test Functions

Evaluating the Results

Potential Causes of Pressure Test Failure

After isolating and fixing the cause of the leak, repeat the pressure test to
confirm the system is pressure tight.

Potential Cause (Pump)

Corrective Action

Purge valve open.

Close the purge valve.

Loose or leaky fitting.

Tighten the fitting or exchange the capillary.

Damaged pump seals or plungers.

Run the leak test to confirm the leak.

Loose purge valve.

Tighten the purge valve nut (14 mm wrench).

Potential Cause (Autosampler)

Corrective Action

Loose or leaky fitting.

Tighten or exchange the fitting or capillary.

Rotor seal (injection valve).

Exchange the rotor seal.

Damaged metering seal or plunger.

Exchange the metering seal. Check the
plunger for scratches. Exchange the plunger if
required.

Needle seat.

Exchange the needle seat.

Potential Cause (Column Compartment)

Corrective Action

Loose or leaky fitting.

Tighten or exchange the fitting or capillary.

Rotor seal (column switching valve).

Exchange the rotor seal.