Assembling the main cover – Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 155

Advertising
background image

155

Repairing the Pump

Assembling the Main Cover

Assembling the Main Cover

N O T E

The cover kit contains all parts, but it is not assembled.

WA R NI N G

In case you insert the left or right side in the opposite position, you
may not be able to remove the side from the top part.

When required

If cover is broken

Tools required

None

Parts required

Cover kit 5062-8565 (includes base, top, left and right)

1

Place the top part on the bench and insert the
left and right side into the top part.

2

Replace the cover.

3

Replace the quaternary pump into the stack and reconnect the cables and capillaries.

4

Turn on the quaternary pump.

Front