Introduction to the Quaternary Pump

Setting the 8-bit Configuration Switch

Switches 1 and 2 do not force storage of this set of parameters in non-volatile
memory. Returning the switches 1 and 2 to other positions (other than being
both up) will allow for normal operation.

If you use the following switch settings and power the instrument up again,
the instrument firmware stays in the resident part, that is, it is not operable as
a quaternary pump. It only uses basic functions of the operating system, for
example, for communication, and so on.

To return to normal operation, set switches back to your GPIB or RS 232C
configuration settings.

Table 42

Stay Resident Settings

Mode Select 1

2

3

4

5

6

7

8

TEST/BOOT

1

1

0

0

1

0

0

0