Forced cold-start settings, Stay resident settings – Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 226

Advertising
background image

226

Introduction to the Quaternary Pump

Setting the 8-bit Configuration Switch

Forced Cold-Start Settings

Switches 1 and 2 do not force storage of this set of parameters in non-volatile
memory. Returning the switches 1 and 2 to other positions (other than being
both up) will allow for normal operation.

C A U T I O N

A forced cold start erases all methods and data stored in non-volatile memory.
Exceptions are diagnose and repair logbooks will not be

erased.

If you use the following switch settings and power the instrument up again, a
forced cold start has been completed.

To return to normal operation, set switches back to your GPIB or RS 232C
configuration settings.

Stay Resident Settings

Firmware update procedures may require this mode in case of firmware
loading errors.

Table 40

Parity Settings

Switches

Parity

7

8

0

0

No Parity

0

1

Odd Parity

1

0

Even Parity

Table 41

Forced Cold Start Settings

Mode Select 1

2

3

4

5

6

7

8

TEST/BOOT

1

1

0

0

0

0

0

1