Installing the Pump

Optimizing the Stack Configuration

Optimizing the Stack Configuration

If your quaternary pump is part of a complete 1100 Series system, you can
ensure optimum performance by limiting the configuration of the system
stack to the following configuration. This configuration optimizes the system
flow path, ensuring minimum delay volume.

Figure 1

Recommended Stack Configuration (Front View)

N O T E

For a detailed view of the flow connections refer to the section “Flow
Connections” in chapter 1 of the reference manuals of the individual modules.

Autosampler

Quaternary pump

Vacuum
degasser

Solvent
cabinet

Detector

Column compartment

Control Module

Flow connections in the stack:
Example setup with 0.17mm ID green capillaries

Solvent bottles - degasser:
G1311-60003 (bottle-head assembly, PTFE-tubings)

Degasser - pump:
G1322-67300 (PTFE-tubings)

Pump - autosampler:
G1312-67305 (SST, green)
Pump purge valve - waste:
5062-2461 (PTFE tubing wide bore, reorder pack)

Autosampler - column compartment:
G1313-87305 (SST, green)

Detector - waste:
DAD 0890-1713 (PTFE, wide bore)
VWD 5062-8535 (PEEK)
5062-2463 (corrugated waste tubing, reorder pack)

Column - detector:
DAD G1315-87311 (SST, coated)
VWD 5062-8522 (PEEK)

Column compartment - column:
G1316-87300 (SST, green)