Agilent Technologies Quaternary Pump 1100 Series User Manual

Page 16

Advertising
background image

16

Installing the Pump

Site Requirements

Table 1

Physical Specifications

Type

Specification

Comments

Weight

11 kg (25 lbs)

Dimensions
(height × weight × depth)

140 Ч 345 Ч 435 mm
(5.5 Ч 13.5 Ч 17 inches)

Line voltage

100–120 or 220–240 VAC,

±

10%

Wide-ranging capability

Line frequency

50 or 60 Hz,

±

5%

Power consumption

220 VA

Maximum

Ambient operating temperature

4–55

°

C (41–131

°

F)

Ambient non-operating temperature

-40–70

°

C (-4–158

°

F)

Humidity

< 95%, at 25–40

°

C (77–104

°

F)

Non-condensing

Operating Altitude

Up to 2000 m (6500 ft)

Non-operating altitude

Up to 4600 m (14950 ft)

For storing the quaternary
pump

Safety standards: IEC, CSA, UL

Installation Category II, Pollution Degree 2

Advertising