Introduction to the Quaternary Pump

Setting the 8-bit Configuration Switch

Setting the 8-bit Configuration Switch

The 8-bit configuration switch is located next to the GPIB connector. Switch
settings provide configuration parameters for GPIB address, serial
communication protocol and instrument specific initialization procedures.

Figure 50

8-bit Configuration Switch

.

Switches 1 and 2 define which set of parameters (for example, for GPIB,
RS-232C, and so on) will be changed. Once the change has been completed,
the instrument must be powered up again in order to store the values in the
non-volatile memory.

factory setting is shown
for quaternary pump

0

1

Table 35

8-bit Configuration Switch

Mode Select 1

2

3

4

5

6

7

8

GPIB

0

0

GPIB Address

RS-232C

0

1

Baudrate

Data Bits Parity

Reserved

1

0

Reserved

TEST/BOOT

1

1

RSVD

SYS

RSVD

RSVD

FC