Parts and Materials

Remote Cables

Agilent 1100 to 3396A Integrators

Agilent 1100 to 3396 Series II / 3395A Integrators

Use the cable 03394-60600 and cut pin #5 on the integrator side. Otherwise
the integrator prints START; not ready.

Connector
03394-60600

Pin
3394

Pin
Agilent 1100

Signal Name

Active
(TTL)

9

1 - White

Digital ground

NC

2 - Brown

Prepare run

Low

3

3 - Gray

Start

Low

NC

4 - Blue

Shut down

Low

NC

5 - Pink

Not connected

NC

6 - Yellow

Power on

High

5,14

7 - Red

Ready

High

1

8 - Green

Stop

Low

NC

9 - Black

Start request

Low

13, 15

Not connected