PR 5220 Instrument Manual

Commissioning

Sartorius

EN-103

Example:

The output signal (Limit 1 out) of limit 1 switches
OFF above a weight of 900 kg. The output signal
(Limit 2 out) of limit 2 switches OFF below a
weight of 290 kg. Both limit values have a
hysteresis of 10 kg.
In the event of a power failure, the two outputs
go to OFF, thus indicating underfill and overfill at
the same time.

If the limits (Limit 1 and Limit 2) for 'On' and 'Off'
are equal (on = off), output 1 (Limit 1 out)
switches ON, when the weight (Wgt) exceeds the
value and output 2 (Limit 2 out) switches OFF,
when the weight drops below the value.

1. Determining an action
Determine the action for the rising edge of the reference signal under [Limit 1 on] from the following list
(here: Marker 1 is set when 900 kg are exceeded).
Accordingly, an action for [Limit 1 off] can be determined.

Setup/

Limit parameter

Limit 1 on

900 kg

Action

set marker 1

X64=1

Condition

no condition

-----

Function

SPM Bit

-no action-

-----

no function

set marker 1

X64=1

Set marker 1

set marker 2

X65=1

Set marker 2

set marker 3

X66=1

Set marker 3

clr marker 1

X64=0

Clear marker 1

clr marker 2

X65=0

Clear marker 2

clr marker 3

X66=0

Clear marker 3


Note:

The limit values can be assigned to the outputs directly in the I/O parameters.